Test method
Delivered energy on seven loads (25 to 175 Ω) with ±3 J or ±15 % tolerance
Free material · Defibrillator and AED
Execution guide to IEC 60601-2-4. The standard declares three essential performances: deliver energy, deliver synchronized and discriminate shockable from non-shockable rhythm. A report that only logs energy proves one third.
Free. For clinical engineering, biomedical service and emergency fleets.
No spam. We only use your data for this request.
“The technical basis connects delivered energy on seven loads (25 to 175 ω) with ±3 j or ±15 % tolerance with charge time: why frequent and infrequent use have different limits, so the material supports an auditable decision instead of a generic checklist.”
Who uses ArkmedsHospitals, networks and clinical engineering providers in Brazil and LatAm
Essential details, in under a minute.
We send access to your email.
Use it in daily clinical engineering, with support if needed.
Delivered energy on seven loads (25 to 175 Ω) with ±3 J or ±15 % tolerance
Charge time: why frequent and infrequent use have different limits
Execution sheet with load, declared and measured values on each line
Execution guide to IEC 60601-2-4. The standard declares three essential performances: deliver energy, deliver synchronized and discriminate shockable from non-shockable rhythm. A report that only logs energy proves one third.
Use it as field support. Your manufacturer instructions, applicable standards and local approved procedure always prevail.