Test method
Five points on a plane 10 cm above the mattress, where the baby is
Free material · Neonatal incubator
Execution guide to IEC 60601-2-19. An incubator with a correct center and a corner 1.5 °C colder passes any single-point test, and still cools the neonate placed in that corner.
Free. For clinical engineering and teams supporting neonatology.
No spam. We only use your data for this request.
“The technical basis connects five points on a plane 10 cm above the mattress, where the baby is with air velocity, noise and overshoot, which almost no report includes, so the material supports an auditable decision instead of a generic checklist.”
Who uses ArkmedsHospitals, networks and clinical engineering providers in Brazil and LatAm
Essential details, in under a minute.
We send access to your email.
Use it in daily clinical engineering, with support if needed.
Five points on a plane 10 cm above the mattress, where the baby is
Air velocity, noise and overshoot, which almost no report includes
Execution sheet with the five values and the mean on each line
Execution guide to IEC 60601-2-19. An incubator with a correct center and a corner 1.5 °C colder passes any single-point test, and still cools the neonate placed in that corner.
Use it as field support. Your manufacturer instructions, applicable standards and local approved procedure always prevail.